RecallRadar

FDA Device recall Z-2156-2026

Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I

On 2026-05-20 the FDA classified a Class I recall of Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I by Arrow International, citing 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2156-2026
ClassificationClass I
Statusongoing
Initiated2026-04-10
Published2026-05-20
Last updated2026-09-13 11:58 UTC
CompanyArrow International
DistributionUS

Product

Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I

Reason

16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.

Identifiers

code_infoREF/UDI-DI/Lot: CS-15242-I/ 30801902195280/33F23G0536; CS-15282-I/30801902195532/33F23C0877, 33F24A0281, 33F24H0041; CS-15322-I/30801902197451/33F23G0545

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2156-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.