FDA Device recall Z-2155-2018
MEDLINE ADULT MANUAL RESUSCITATOR WITH LARGE ADULT MASK, TUBE RESERVOIR, FILTER AND 7 FT. OXYGEN TUBING, REF CPRM1126FL
- FDA Device
- Class II
- terminated
- Published 2018-06-20
On 2018-06-20 the FDA classified a Class II recall of MEDLINE ADULT MANUAL RESUSCITATOR WITH LARGE ADULT MASK, TUBE RESERVOIR, FILTER AND 7 FT. OXYGEN TUBING, REF CPRM1126FL by Sun Med, citing there is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2155-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-05-24 |
| Published | 2018-06-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Sun Med |
| Distribution | IL, OH |
Product
MEDLINE ADULT MANUAL RESUSCITATOR WITH LARGE ADULT MASK, TUBE RESERVOIR, FILTER AND 7 FT. OXYGEN TUBING, REF CPRM1126FL
Reason
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
Identifiers
| code_info | Lot Numbers: 313363 & 313738 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.