FDA Device recall Z-2152-2023
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
- FDA Device
- Class I
- ongoing
- Published 2023-07-26
On 2023-07-26 the FDA classified a Class I recall of Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85 by Datascope, citing an unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected inter….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2152-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-06-05 |
| Published | 2023-07-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Datascope |
| Distribution | US |
Product
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
Reason
An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.
Identifiers
| code_info | All Unit Serial Numbers Model UDI 0998-00-0800-75 10607567112312 0998-00-0800-83 10607567108407 0998-00-0800-85 10607567113449 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2152-2023%22