FDA Device recall Z-2149-2023
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0…
- FDA Device
- Class I
- ongoing
- Published 2023-07-26
On 2023-07-26 the FDA classified a Class I recall of Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0… by Datascope, citing IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low ba….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2149-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-06-05 |
| Published | 2023-07-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Datascope |
| Distribution | US |
Product
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution
Reason
IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.
Identifiers
| code_info | All Unit Serial Numbers Model UDI 0998-00-0800-31 10607567109053 0998-00-0800-32 10607567111117 099…All Unit Serial Numbers Model UDI 0998-00-0800-31 10607567109053 0998-00-0800-32 10607567111117 0998-00-0800-33 10607567109008 0998-00-0800-34 10607567111940 0998-00-0800-35 10607567109107 0998-00-0800-45 10607567108421 0998-00-0800-52 10607567108438 0998-00-0800-53 10607567108391 0998-00-0800-55 10607567108414 0998-00-0800-65 10607567113432 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2149-2023%22