RecallRadar

FDA Device recall Z-2147-2016

AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae…

On 2016-07-20 the FDA classified a Class II recall of AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae… by Sorin Group Usa, citing sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2147-2016
ClassificationClass II
Statusterminated
Initiated2016-06-22
Published2016-07-20
Last updated2026-09-13 11:54 UTC
CompanySorin Group Usa
DistributionUS

Product

AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.

Reason

Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.

Identifiers

code_info
Lots: 1505600017 ; 1505600018 ; 1506500005 ; 1509200078 ; 1513100112 ; 1513900027 ; 1513900028 ; 1514700045 ; 151660019…Lots: 1505600017 ; 1505600018 ; 1506500005 ; 1509200078 ; 1513100112 ; 1513900027 ; 1513900028 ; 1514700045 ; 1516600192 ; 1516800126 ; 1517800004 ; 1519500037 ; 1520100049 ; 1520100050 ; 1520900031 ; 1520900033 ; 1521600126 ; 1523000093 ; 1523600120 ; 1525100023 ; 1525800026 ; 1526500032 ; 1526500033 ; 1526500034 ; 1526500072 ; 1527200050 ; 1527200051 ; 1527200052 ; 1527800163 ; 1529300026 ; 1529300027 ; 1529300028 ; 1533500033 ; 1534200034 ; 1601900027 ; 1601900028 ;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2147-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.