RecallRadar

FDA Device recall Z-2143-2023

HeartSine, REF: PAD-PAK-01, PAD-PAK-03 AND PAD-PAK-03j, CE0123 sold separately and with the HeartSine Samaritan Public Access Defibrillator…

On 2023-07-19 the FDA classified a Class II recall of HeartSine, REF: PAD-PAK-01, PAD-PAK-03 AND PAD-PAK-03j, CE0123 sold separately and with the HeartSine Samaritan Public Access Defibrillator… by Heartsine Technologies, citing single use battery and electrode cartridge may be rendered inoperable due to depleted battery cells which could prevent the defibrillator from analyzing patient condition or deliv….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2143-2023
ClassificationClass II
Statusongoing
Initiated2023-04-26
Published2023-07-19
Last updated2026-09-13 11:56 UTC
CompanyHeartsine Technologies
DistributionUS

Product

HeartSine, REF: PAD-PAK-01, PAD-PAK-03 AND PAD-PAK-03j, CE0123 sold separately and with the HeartSine Samaritan Public Access Defibrillator models: 350P/360P/450P/500P

Reason

Single use battery and electrode cartridge may be rendered inoperable due to depleted battery cells which could prevent the defibrillator from analyzing patient condition or delivering therapy correctly.

Identifiers

code_info
Model #/UDI-DI #: 360-BAS-UK-10/05060167122606, 350-BAS-UK-10/05060167121418, 350-BAS-AS-10/05060167121371, 350-BAS-CF-…Model #/UDI-DI #: 360-BAS-UK-10/05060167122606, 350-BAS-UK-10/05060167121418, 350-BAS-AS-10/05060167121371, 350-BAS-CF-10/05060167124747, 350-BAS-CN-10/05060167121593, 350-BAS-JA-08/05060167124525, 350-BAS-KO-10/05060167121517, 350-BAS-MS-10/05060167120800, 350-BAS-USROW-10/05060167121630, 360-BAS-AS-10/05060167122514, 360-BAS-CN-10/05060167125751, 360-BAS-KO-10/05060167122712, 360-BAS-SJ-10/05060167128028, 450-BAS-JA-08/05060167127663, 500-BAS-AS-10/05060167125812, 500-BAS-CF-10/05060167125836, 500-BAS-CN-10/05060167125843, 500-BAS-KO-10/05060167126444, 500-BAS-TH-10/05060167125997, 500-BAS-UK-10/05060167122453, PAD-PAK-01/05060167120466,PAD-PAK-03/05060167121234,PAD-PAK-03j/05060167124754 Models/Lot # 360-BAS-UK-10/ A3659, A3655, A3664, A3654, A3803, A3829, A3840, A3842, A3848, A3840, A3843, A3842, A3844, A3842, J0754, J0754, J0755, J0752, J0798; 350-BAS-UK-10/ A3664, A3799, A3803, A3829,A3840,A3846, J0753, J0753, J0754, J0756; 350-BAS-AS-10/ A3656; 350-BAS-CF-10/ A3635, A380

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2143-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.