FDA Device recall Z-2143-2020
Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Out…
- FDA Device
- Class II
- terminated
- Published 2020-06-03
On 2020-06-03 the FDA classified a Class II recall of Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Out… by Spectranetics, citing there is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations. All users, such as those performing lead extraction cases may not….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2143-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-03-31 |
| Published | 2020-06-03 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Spectranetics |
| Distribution | US |
Product
Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Outer Diameter 31 mm, Balloon Length 80 mmm, Rx ONLY, CE Marking, Sterile EO, UDI: 00813132024666
Reason
There is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations. All users, such as those performing lead extraction cases may not be aware of this risk. It was determined that current labeling is insufficient to ensure all potential users minimize the time the balloon remains in vasculature. If left in too long it could lead to embolization.
Identifiers
| code_info | All lots |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Other recalls by Spectranetics
- Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm,…
FDA Device · Class I · 2025-02-26 · Spectranetics - Turbo Elite Laser Atherectomy Catheter Model Number / UDI-DI code / Product Description 414-151 / 989930000531 / 1.4 mm OTW 417-152 / 98993…
FDA Device · Class II · 2024-12-25 · Spectranetics - Spectranetics Turbo-Elite Laser Atherectomy Catheter 2.0 RX REF 420-159 Spectranetics Turbo-Elite Laser Atherectomy Catheter 2.0 OTW REF 42…
FDA Device · Class II · 2023-12-20 · Spectranetics - Philips Laser System, REF: LAS-100
FDA Device · Class II · 2022-11-23 · Spectranetics - Philips Laser System, REF LAS-100, Intermittent Operation, Duty cycle dependent upon device in use 100 - 240 VAC, 50/60HZ, 16 AM
FDA Device · Class II · 2022-07-20 · Spectranetics
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2143-2020%22