RecallRadar

FDA Device recall Z-2143-2020

Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Out…

On 2020-06-03 the FDA classified a Class II recall of Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Out… by Spectranetics, citing there is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations. All users, such as those performing lead extraction cases may not….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2143-2020
ClassificationClass II
Statusterminated
Initiated2020-03-31
Published2020-06-03
Last updated2026-09-13 11:55 UTC
CompanySpectranetics
DistributionUS

Product

Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Outer Diameter 31 mm, Balloon Length 80 mmm, Rx ONLY, CE Marking, Sterile EO, UDI: 00813132024666

Reason

There is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations. All users, such as those performing lead extraction cases may not be aware of this risk. It was determined that current labeling is insufficient to ensure all potential users minimize the time the balloon remains in vasculature. If left in too long it could lead to embolization.