FDA Device recall Z-2141-2020
Accelerator a3600 Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N F…
- FDA Device
- Class II
- terminated
- Published 2020-06-03
On 2020-06-03 the FDA classified a Class II recall of Accelerator a3600 Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N F… by Inpeco, citing the identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Au….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2141-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-04-07 |
| Published | 2020-06-03 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Inpeco |
| Distribution | US |
Product
Accelerator a3600 Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Reason
The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.
Identifiers
| code_info | Serial Numbers: ACP.290 ACP.343 ACP.365 |
|---|
Status history