AUTOMIX Plus High Speed Compounder System, product 2M8075
- FDA Device
- Class I
- terminated
- Published 2012-08-15
On 2012-08-15 the FDA classified a Class I recall of AUTOMIX Plus High Speed Compounder System, product 2M8075 by Baxter Healthcare, citing after fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press.….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2140-2012 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2012-07-13 |
| Published | 2012-08-15 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Baxter Healthcare |
| Distribution | US |
Product
AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.
Reason
After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
Identifiers
| code_info | Product code 2M8075, all serial numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2140-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.