RecallRadar

FDA Device recall Z-2138-2012

TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC, 132 Citizens Blvd., Simpsonville, KY 40067

On 2012-08-15 the FDA classified a Class II recall of TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC, 132 Citizens Blvd., Simpsonville, KY 40067 by Tango3, citing during an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2138-2012
ClassificationClass II
Statusterminated
Initiated2012-03-26
Published2012-08-15
Last updated2026-09-13 11:53 UTC
CompanyTango3
DistributionCA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, ME, NH, NJ, NM, OH, PA, RI, TX, VA

Product

TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC, 132 Citizens Blvd., Simpsonville, KY 40067. Models are WOZ-250-PE, WOZ-240-SS, WOZ-180-SS. The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility.

Reason

During an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either not reviewed or received proper clearance from FDA. There was no assurance that all systems in the field were working properly and operating at a minimum of 0.2ppm of ozone.

Identifiers

code_info
Model WOZ -250-PE, Serial #s: 1031, 1032, 1033, 1035, 1036, 1038, 1039, 1042, 1043, 1044, 1045, 1046, 1047, 1048, 1049,…Model WOZ -250-PE, Serial #s: 1031, 1032, 1033, 1035, 1036, 1038, 1039, 1042, 1043, 1044, 1045, 1046, 1047, 1048, 1049, 1050, 1051, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1063, 1065, 1066, 1067, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076 & 1077. Model WOZ-180-SS, Serial #s: 1034, 1037, 1040 & 1064. Model WOZ-240-SS, Serial #s: 1041 & 1052.

Status history