RecallRadar

FDA Device recall Z-2136-2021

Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat inf…

On 2021-08-04 the FDA classified a Class II recall of Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat inf… by Medtronic Vascular, citing during product complaint investigation, it was determined that during stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2136-2021
ClassificationClass II
Statusongoing
Initiated2021-06-07
Published2021-08-04
Last updated2026-09-13 11:56 UTC
CompanyMedtronic Vascular
DistributionGA, MA, MD, NY, SC, TX, WA, WV

Product

Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Reason

During product complaint investigation, it was determined that during stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.

Identifiers

code_info
Endurant IIs Stent Graft System: Model (REF):ESBF3214C103E GTIN: 00643169439962 Serial Numbers: V29780773; V29780771; V…Endurant IIs Stent Graft System: Model (REF):ESBF3214C103E GTIN: 00643169439962 Serial Numbers: V29780773; V29780771; V29780759; Model (REF): ESBF3614C103EE GTIN: 00643169780460 Serial Numbers: V29781932; V29781933; V29781934; V29781935; V29781936; V29781937; V29781938; V29781939; V29781940; V29781941; V29781942; V29781943; V29781944; V29781945; V29781946; V29781947; V29781948; V29781951;