ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determinatio…

On 2020-06-03 the FDA classified a Class II recall of ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determinatio… by Beckman Coulter, citing electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associat….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2136-2020
ClassificationClass II
Statusterminated
Initiated2020-01-29
Published2020-06-03
Last updated2026-09-13 11:55 UTC
CompanyBeckman Coulter
DistributionUS

Product

ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.

Reason

Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.

Identifiers

code_infoLot Number: M909206