FDA Device recall Z-2136-2018
cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vit…
- FDA Device
- Class II
- terminated
- Published 2018-06-20
On 2018-06-20 the FDA classified a Class II recall of cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vit… by Roche Diagnostics, citing the syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2136-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-04-03 |
| Published | 2018-06-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Roche Diagnostics |
| Distribution | CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, VA |
Product
cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement.
Reason
The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.
Identifiers
| code_info | Serial numbers: 1744-08 1744-09 1744-10 1758-08 1606-09 1752-05 1752-06 1630-08 1738-01 1744-07 1758-09 1759…Serial numbers: 1744-08 1744-09 1744-10 1758-08 1606-09 1752-05 1752-06 1630-08 1738-01 1744-07 1758-09 1759-04 1758-06 1758-07 1743-09 1743-10 1752-04 1759-05 17C5-06 17C5-07 1615-08 1612-10 17C5-09 1744-05 1744-06 1743-07 1743-08 1744-03 1744-04 1759-06 17C5-05 17C5-08 17C5-10 17C6-01 17C6-02 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2136-2018%22