FDA Device recall Z-2135-2012
RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination…
- FDA Device
- Class II
- terminated
- Published 2012-08-08
On 2012-08-08 the FDA classified a Class II recall of RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination… by Stryker Instruments Div, citing stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the act….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2135-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-07-17 |
| Published | 2012-08-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Instruments Div |
| Distribution | US |
Product
RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.
Reason
Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed
Identifiers
| code_info | REF 0406-630-125, lot 1000037535 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2135-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.