FDA Device recall Z-2134-2014
Patient Data Manager 2.0 (Content manager 2.0, Patient Browser 4.0, DICOM Viewer 2.0) Brainlab Digital Lightbox
- FDA Device
- Class II
- terminated
- Published 2014-08-06
On 2014-08-06 the FDA classified a Class II recall of Patient Data Manager 2.0 (Content manager 2.0, Patient Browser 4.0, DICOM Viewer 2.0) Brainlab Digital Lightbox by Brainlab Ag, citing the following defects have been identified for data loaded with Patient Data Manager version 2.0 (including subversions 2.0.0, 2.0.1 and 2.0.2): 1) When loading non-square pixel i….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2134-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-06-27 |
| Published | 2014-08-06 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Brainlab Ag |
| Distribution | US |
Product
Patient Data Manager 2.0 (Content manager 2.0, Patient Browser 4.0, DICOM Viewer 2.0) Brainlab Digital Lightbox. A system for the display of medical images. Model numbers 25100-05 and 25100-06. The software can transfer images to and from picture archiving and communication systems (PACS), file servers, or removable storage media.
Reason
The following defects have been identified for data loaded with Patient Data Manager version 2.0 (including subversions 2.0.0, 2.0.1 and 2.0.2): 1) When loading non-square pixel image data with PDM 2.0, the data set is not correctly transferred and displayed in the corresponding Brainlab navigation or planning software. When a data set is loaded, the pixels of the data set are sorted in order to
Identifiers
| code_info | Model/Catalog Numbers: 25100-05 PATIENT DATA MANAGER 2.0.11 25100-06 PATIENT DATA MANAGER 2.0.2. Software Versions: 2.0.0, 2.0.1, and 2.0.2 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Brainlab Ag
- Origin Data Management software versions 3.1.0, 3.1.1, 3.1.2, 3.2.0, 3.2.1
FDA Device · Class II · 2025-03-26 · Brainlab Ag - Cirq Arm System 2.0
FDA Device · Class II · 2023-04-05 · Brainlab Ag - Brainlab ExacTrac Dynamic software, Medical Charged-Particle Radiation Therapy System, model 20910-01E and 20910-01F
FDA Device · Class II · 2023-03-22 · Brainlab Ag - ExacTrac Dynamic software model numbers: 20910-01, ETD Positioning and Moni. 1.0.0 INST
FDA Device · Class II · 2022-03-16 · Brainlab Ag - ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at…
FDA Device · Class II · 2021-06-16 · Brainlab Ag