RecallRadar

FDA Device recall Z-2132-2016

Item # 4042-CPR Rescue Kit, containing CPR shield with one way valve & barrier filter. brand Lifeline, UPC 845174004429 each

On 2016-07-13 the FDA classified a Class II recall of Item # 4042-CPR Rescue Kit, containing CPR shield with one way valve & barrier filter. brand Lifeline, UPC 845174004429 each by Lifeline First Aid, citing first aid kits or rescue kits are recalled because they contains the CPR Shield with One Way Valve and Barrier Filter (mask) and this mask is cleared only for prescription use and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2132-2016
ClassificationClass II
Statusterminated
Initiated2016-05-16
Published2016-07-13
Last updated2026-09-13 11:54 UTC
CompanyLifeline First Aid
DistributionAZ, CA, CO, FL, GA, IL, KY, MD, MN, MO, MT, NE, NJ, NY, OK, OR, PA, SC, TN, UT, VA, WA, WI

Product

Item # 4042-CPR Rescue Kit, containing CPR shield with one way valve & barrier filter. brand Lifeline, UPC 845174004429 each; UPC 845174004436 master case of 6. Item #4043-CPR Rescue Kit, containing CPR shield with one way valve & barrier filter. brand Lifeline. UPC Red color kit: 845174005464; UPC Lime color kit: 845174005594; UPC Purple color: 845174005600; UPC of master case of 6 845174005471 (2 each color) Item # 4037, Outfitter First Aid Kit, CPR shield with one way valve & barrier filter, brand Lifeline, UPC 845174003453

Reason

First aid kits or rescue kits are recalled because they contains the CPR Shield with One Way Valve and Barrier Filter (mask) and this mask is cleared only for prescription use and is not cleared for over-the-counter (OTC) use.

Identifiers

upc845174004429
upc845174004436
upc845174003453
code_info
item #4042 and item #4043 have the expiration date: 11/30/2017. Item #4037 has the following lots and expiration dat…item #4042 and item #4043 have the expiration date: 11/30/2017. Item #4037 has the following lots and expiration dates: Lot # Expiration Date 101483; 3/15 101546; 6/15 101682; 7/15 101804; 1/16 101812; 1/16 102032; 5/16 102109; 5/16 102098; 5/17 102130; 5/16 102296; 10/16 102317; 10/16 102295; 10/16 102372; 1/17 102638; 1/17 102694; 8/17

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2132-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.