RecallRadar

FDA Device recall Z-2131-2014

Zimmer Periarticular Locking Plate System Distal Radial Volar Plate Lateral Column (1st generation stainless steel only)

On 2014-08-06 the FDA classified a Class II recall of Zimmer Periarticular Locking Plate System Distal Radial Volar Plate Lateral Column (1st generation stainless steel only) by Zimmer, citing A complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pu….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2131-2014
ClassificationClass II
Statusterminated
Initiated2014-06-10
Published2014-08-06
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionCA, OH

Product

Zimmer Periarticular Locking Plate System Distal Radial Volar Plate Lateral Column (1st generation stainless steel only). Orthopedic fixation plate. For temporary internal fixation and stabilization of osteotomies and fractures.

Reason

A complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pulling through plate hole.

Identifiers

code_infoPart Number 00-2358-014-04; lot 62276042

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2131-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.