FDA Device recall Z-2129-2018
Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000…
- FDA Device
- Class II
- terminated
- Published 2018-06-13
On 2018-06-13 the FDA classified a Class II recall of Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000… by Orthofix, citing firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws wer….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2129-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-04-06 |
| Published | 2018-06-13 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Orthofix |
| Distribution | CA, DC, FL, GA, IL, IN, LA, MD, MI, MN, NC, NJ, NV, OH, OK, PA, SC, TX, VA |
Product
Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000L); b) Small/0MM Step/Right (002-C-01000R); c) Large/0MM Step/Left (002-C-02000L); d) Large/0MM Step/Right (002-C-02000R); e) Small/2MM Step/Left (002-C-01002L); f) Small/2MM Step/Right (002-C-01002R); g) Large/2MM Step/Left (002-C-02002L); h) Large/2MM Step/Right (002-C-02002R)
Reason
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Identifiers
| code_info | (UDI)/Lot: a) (18056099647588) B1096741; b) (18056099647595) B1117260, B1099064; c) (18056099647625) B1117352, B109…(UDI)/Lot: a) (18056099647588) B1096741; b) (18056099647595) B1117260, B1099064; c) (18056099647625) B1117352, B1096743; d) (18056099647632) B1099065 e) (18056099647601) B1117265, f) (18056099647618) B1098803, B1116361; g) 18056099647649) B1200565, B1116364; h) (18056099647656) B1123046, |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2129-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.