FDA Device recall Z-2120-2012
Octane-C Cervical, 6¿ Lordotic Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7)
- FDA Device
- Class II
- terminated
- Published 2012-08-08
On 2012-08-08 the FDA classified a Class II recall of Octane-C Cervical, 6¿ Lordotic Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7) by Exactech, citing exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Inte….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2120-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-06-15 |
| Published | 2012-08-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Exactech |
| Distribution | AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, TN, TX, VA, WA |
Product
Octane-C Cervical, 6¿ Lordotic Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7).
Reason
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year
Identifiers
| code_info | Sizes: 5 mm Catalog # 05-010-02-0605 All serial numbers 6 mm Catalog # 05-010-02-0606 All serial numbers 7 mm Catal…Sizes: 5 mm Catalog # 05-010-02-0605 All serial numbers 6 mm Catalog # 05-010-02-0606 All serial numbers 7 mm Catalog # 05-010-02-0607 All serial numbers 8 mm Catalog # 05-010-02-0608 All serial numbers 9 mm Catalog # 05-010-02-0609 All serial numbers 10mm Catalog # 05-010-02-0610 All serial numbers 11mm Catalog # 05-010-02-0611 All serial numbers 12mm Catalog # 05-010-02-0612 All serial numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2120-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.