FDA Device recall Z-2118-2013

Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage…

On 2013-09-04 the FDA classified a Class II recall of Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage… by Ventana Medical Systems, citing ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the u….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2118-2013
ClassificationClass II
Statusterminated
Initiated2008-06-12
Published2013-09-04
Last updated2026-09-13 11:53 UTC
CompanyVentana Medical Systems
DistributionUS

Product

Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The Benchmark product lines are intended to automatically stain histological or cytological specimens on microscope slides with specific immunohistochemistry or in situ hybridization reagents for in vitro diagnostic use. The Discovery Systems are research use only instruments with similar functionality.

Reason

Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.

Identifiers

code_infoBenchMark XT Instruments-S/N: 711869-712465; BenchMark LT instruments-S/N: 610321-610345. Discovery XT Instruments-S/N: 610321-610345.