RecallRadar

FDA Device recall Z-2116-2018

Rival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-011C

On 2018-06-13 the FDA classified a Class II recall of Rival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-011C by Orthofix, citing firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws wer….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2116-2018
ClassificationClass II
Statusterminated
Initiated2018-04-06
Published2018-06-13
Last updated2026-09-13 11:54 UTC
CompanyOrthofix
DistributionCA, DC, FL, GA, IL, IN, LA, MD, MI, MN, NC, NJ, NV, OH, OK, PA, SC, TX, VA

Product

Rival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-011C

Reason

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Identifiers

code_infoUDI: 18056099648592, B1115657, B1107222, B1107223

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.