RecallRadar

FDA Device recall Z-2114-2020

FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interf…

On 2020-06-03 the FDA classified a Class II recall of FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interf… by Inpeco, citing the sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2114-2020
ClassificationClass II
Statusterminated
Initiated2020-04-07
Published2020-06-03
Last updated2026-09-13 11:55 UTC
CompanyInpeco
DistributionIL, NY

Product

FlexLab Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output Module ROM400 (P/N FLX-289), XN-9000 Interface Module (P/N FLX-290), Advia 2120 LAS Interface Module (P/N FLX-219), Vertical Transportation Module (P/N FLX -204), Rack Input Module (P/N FLX-214), AUWi System Interface Module (P/N FLX-286), Track To Rack Interface Module (P/N FLX-295) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Reason

The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.

Identifiers

code_infoSerial Numbers: FLX.0133 FLX.0191 FLX.0192 FLX.0197 FLX.0198 FLX.0201 FLX.0202 FLX.027 FLX.0200 FLX.0181

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.