FDA Device recall Z-2113-2013
DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx On…
- FDA Device
- Class II
- terminated
- Published 2013-09-04
On 2013-09-04 the FDA classified a Class II recall of DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx On… by Deroyal Industries, citing deRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing whic….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2113-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-07-15 |
| Published | 2013-09-04 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Deroyal Industries |
| Distribution | US |
Product
DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen.
Reason
DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.
Identifiers
| code_info | All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214;…All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214; REF 28-0214 Hi-Flow Insufflation Tubing Lot #3194170; REF 28-0216 Tubing, Insuf w/.3 Micr Filter Lot #29074249; REF 28-0217 Insufflation Tubing, Laparoscopic Lot #32305010; DeRoyal REF 28-0208 Insufflation Tubing w/Vent Lot #29314302; DeRoyal REF 28-0212H Heated Hi-Flow Insufflation Tubing Lot #29000354. |
|---|