FDA Device recall Z-2111-2014
Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to displ…
- FDA Device
- Class II
- terminated
- Published 2014-07-30
On 2014-07-30 the FDA classified a Class II recall of Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to displ… by Siemens Medical Solutions Usa, citing siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2111-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-06-30 |
| Published | 2014-07-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo Imaging XS also supports storage and archiving of Structured DICOM reports. Syngo Imaging XS optionally uses a variety of advanced post processing applications.
Reason
Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge enhancement and Histograms. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d
Identifiers
| code_info | model number: 10496279 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2111-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.