FDA Device recall Z-2100-2012
STERRAD Cyclesure Biological Indicator (BI), P/N 14324 are sold in cases, packaged as part of STERRAD System Validation kits and STERRAD te…
- FDA Device
- Class II
- terminated
- Published 2012-08-08
On 2012-08-08 the FDA classified a Class II recall of STERRAD Cyclesure Biological Indicator (BI), P/N 14324 are sold in cases, packaged as part of STERRAD System Validation kits and STERRAD te… by Advanced Sterilization Products, citing advanced Sterilization Products (ASP) is recalling certain lots of the STERRAD CYCLESURE 24 Biological Indicator (BI) product because it does not have adequate data to support the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2100-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-06-28 |
| Published | 2012-08-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Advanced Sterilization Products |
| Distribution | US |
Product
STERRAD Cyclesure Biological Indicator (BI), P/N 14324 are sold in cases, packaged as part of STERRAD System Validation kits and STERRAD test packs. The STERRAD CYCLESURE 24 Biological Indicator (BI), P/N 14324 is intended to be used as a standard method for frequent monitoring of the STERRAD Sterilization System cycles.
Reason
Advanced Sterilization Products (ASP) is recalling certain lots of the STERRAD CYCLESURE 24 Biological Indicator (BI) product because it does not have adequate data to support the entire duration of the product's labeled shelf-life.
Identifiers
| code_info | Lot# 002127, 003127, 005127, 006127, 007127, 008127, 009127, 010127, 011128, 012127, 013127, 014127, 015127, 016127, 01…Lot# 002127, 003127, 005127, 006127, 007127, 008127, 009127, 010127, 011128, 012127, 013127, 014127, 015127, 016127, 018127, 019127, 020127, 021127, 023127, 024127, 025127, 026127, 027127, 028127, 030127, 031127, 032127, 033127, 034127, 035127, 036127, 037127, 038121, 038127 039127, 040127, 041121, 042127, 043127, 044127, 045121, 045127, 046127, 047127, 048127, 049127, 052127, 053121, 053127, 054127, 055121, 056127, 057127, 058127, 059127, 060127, 061127, 062121, 062127, 063127, 064127, 065127, 067127, 068127, 069121, 069127, 070127, 071127, 072127, 073127, 074127, 076121, 077127, 078127, 078128, 079127, 081127, 082127, 084127, 085127, 086121, 086127, 087127, 089127, 090127, 092127, 093121, 094127, 095127, 100121, 100127, 101127, 102127, 104127, 107121, 107127, 107128, 109127, 109128, 110127, 111127, 114121, 114127, 118127, 121121, 121127, 123127, 124127, 125127, 128121, 128127, 129127, 130127, 131127, 132127, 135121, 135127, 136127, 137127, 138127, 139127, 142121, 142127, 143127, 1441 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2100-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.