FDA Device recall Z-2099-2015

0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover 0702-002-000, Neptu…

On 2015-07-29 the FDA classified a Class II recall of 0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover 0702-002-000, Neptu… by Stryker Instruments Div, citing stryker Instruments is initiating a recall of the Neptune 2 Waste Management System due to complaints related to installation failures of the Neptune 2 diverter valve assembly dur….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2099-2015
ClassificationClass II
Statusterminated
Initiated2015-05-22
Published2015-07-29
Last updated2026-09-13 11:53 UTC
CompanyStryker Instruments Div
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover 0702-002-000, Neptune 2 Rover Ultra (230V) 0702-002-000ES, Neptune 2 Rover (230V) Spanish 0702-002-000FI, Neptune 2 Ultra Rover (230V) Finnish 0702-002-000IT, Neptune 2 Rover (230V) Italian 0702-002-000R, Repair Neptune 2 Rover Ultra (230V) Product Usage: The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor s offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electro cautery or laser devices. Contraindication: connection of the Neptune 2 Waste Management System directly to chest tubes or other closed wound drains. The Stryker Neptune 2 Rover is a mobile unit used to suction, collect fluid waste, and evacuate surgical smoke from a surgical site during operation. Fluid waste is removed from the surgical site through suction tubing connected to inlet ports of disposable manifolds installed in the rover. The Diverter Valve Assembly consists of two diverter valves with three hoses, one inlet and two outlets. The sole purpose of the Diverter Valve Assembly is to direct/divert water from the fresh water inlet to either the large canister or small canister via the pre-fill tank.

Reason

Stryker Instruments is initiating a recall of the Neptune 2 Waste Management System due to complaints related to installation failures of the Neptune 2 diverter valve assembly during docking.

Identifiers

code_infoSerial Numbers ranging from 0803118093 through 1505400223; Distribution or Repair dates: November 1, 2011 to March 31, 2015

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2099-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.