FDA Device recall Z-2098-2014
Sarns Soft-arc Aortic Cannula 8.0mm curved tip, suture flange, 3/8" flare Product Usage: The Aortic Cannula is indicated for use in perfusi…
- FDA Device
- Class II
- terminated
- Published 2014-07-30
On 2014-07-30 the FDA classified a Class II recall of Sarns Soft-arc Aortic Cannula 8.0mm curved tip, suture flange, 3/8" flare Product Usage: The Aortic Cannula is indicated for use in perfusi… by Terumo Cardiovascular Systems, citing during an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2098-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-23 |
| Published | 2014-07-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
Sarns Soft-arc Aortic Cannula 8.0mm curved tip, suture flange, 3/8" flare Product Usage: The Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
Reason
During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications
Identifiers
| code_info | Part No. 16094, Lot #: 655235 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2098-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.