FDA Device recall Z-2097-2015
Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System
- FDA Device
- Class II
- terminated
- Published 2015-07-29
On 2015-07-29 the FDA classified a Class II recall of Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System by Maquet Cardiovascular Us Sales, citing there is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2097-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-06-10 |
| Published | 2015-07-29 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Maquet Cardiovascular Us Sales |
| Distribution | US |
Product
Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.
Reason
There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
Identifiers
| code_info | SN 00048 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2097-2015%22