FDA Device recall Z-2097-2015

Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System

On 2015-07-29 the FDA classified a Class II recall of Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System by Maquet Cardiovascular Us Sales, citing there is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2097-2015
ClassificationClass II
Statusterminated
Initiated2015-06-10
Published2015-07-29
Last updated2026-09-13 11:53 UTC
CompanyMaquet Cardiovascular Us Sales
DistributionUS

Product

Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.

Reason

There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.