RecallRadar

FDA Device recall Z-2095-2026

Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A

On 2026-05-13 the FDA classified a Class II recall of Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A by Medline Industries, citing medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2095-2026
ClassificationClass II
Statusongoing
Initiated2026-01-07
Published2026-05-13
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number:DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number:DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number:DYNDH1934; 6) BREAST BIOPSY PACK, Model Number:DYNJ17489I; 7) C T BIOPSY TRAY-LF, Model Number:DYNJ17558D; 8) RIGHT HEART BIOPSY TRAY, Model Number:DYNJ19063K; 9) VOR PACK-LF, Model Number:DYNJ48539; 10) VOR PACK-LF, Model Number:DYNJ48539A; 11) IR BIOPSY PACK, Model Number:DYNJ59897B; 12) BREAST PACK, Model Number:DYNJ82998; 13) LIVER BIOPSY KIT, Model Number:DYNJ910654

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Identifiers

code_info
1) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBQ267; 2) DYKMBNDL84A, UDI-DI: 10195…1) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBQ267; 2) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBU924; 3) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 23ABK216; 4) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBK913; 5) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBM302; 6) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23ABU170; 7) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBC364; 8) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBL439; 9) DYNDH1143C, UDI-DI: 10193489193336(each), 40193489193337(case), Lot Number: 22GBZ268; 10) DYNDH1873, UDI-DI: 10195327486464(each), 40195327486465(case), Lot Number: 23KBK476; 11) DYNDH1934, UDI-DI: 10195327519728(each), 40195327519729(case), Lot Number: 24FMF173; 12) DYNJ17489

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2095-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.