RecallRadar

FDA Device recall Z-2095-2013

PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT, TT-S s/n 1001-1083

On 2013-09-04 the FDA classified a Class II recall of PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT, TT-S s/n 1001-1083 by Siemens Medical Solutions Usa, citing siemens has become aware of a customer complaint concerning the rail threaded holes of the 550TxT treatment tabletop. Due to a manufacturing error, the screws holding the side rai….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2095-2013
ClassificationClass II
Statusterminated
Initiated2013-07-18
Published2013-09-04
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT, TT-S s/n 1001-1083. Product Usage: The intended use of the SIEMENS branded, ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Reason

Siemens has become aware of a customer complaint concerning the rail threaded holes of the 550TxT treatment tabletop. Due to a manufacturing error, the screws holding the side rails in position on the tabletop might break off.

Identifiers

code_infoONCOR, PRIMUS and ARTISTE Lican systems with tabletop component: Tabletop, 550TxT, TT-A 08632296* Tabletop, 550TxT, TT-S 10046140

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2095-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.