RecallRadar

FDA Device recall Z-2094-2026

Medline Convenience Kits: 1) PICC LINE TRAY, Model Number:00-401993O

On 2026-05-13 the FDA classified a Class II recall of Medline Convenience Kits: 1) PICC LINE TRAY, Model Number:00-401993O by Medline Industries, citing medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2094-2026
ClassificationClass II
Statusongoing
Initiated2026-01-07
Published2026-05-13
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

Medline Convenience Kits: 1) PICC LINE TRAY, Model Number:00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number:ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Number:ACC010268C; 4) TRAY 6 CENTRAL LINE A, Model Number:ACC010487; 5) PEDS PICC INSERTION TRAY, Model Number:CVI3650; 6) LINE ATTIRE KIT, Model Number:DYKM2170; 7) PK BNDL CENT LINE PROTECT KIT, Model Number:DYKM2274; 8) KIT SURG ONC LAP DIAG/PORT INS, Model Number:DYKMBNDL153C; 9) KIT SURG ONC INSERT CENT.VENOU, Model Number:DYKMBNDL43A; 10) KIT SURG ONC PORT INSERTION, Model Number:DYKMBNDL43C; 11) VEIN PROCEDURE KIT, Model Number:DYNDA2076B; 12) VEIN PROCEDURE KIT, Model Number:DYNDA2076D; 13) VEIN PROCEDURE KIT, Model Number:DYNDA2076F; 14) PLACENTA KIT, Model Number:DYNDA2322B; 15) NICU CENTRAL LINE TRAY, Model Number:DYNDC2407A; 16) NICU CENTRAL LINE TRAY, Model Number:DYNDC2407B; 17) PICC LINE TRAY, Model Number:DYNJ40500A; 18) VENOUS ACCESS PACK, Model Number:DYNJ42694C; 19) AV FISTULA CREATION PACK-LF, Model Number:DYNJ53427B; 20) PICC LINE, Model Number:DYNJ55546D; 21) PICC TRAY, Model Number:DYNJ63199A; 22) PICC KIT 4F SL PL MAX BARRIER, Model Number:DYNJ70141MB; 23) PICC KIT 5F DL PL MB, Model Number:DYNJ70152MB; 24) VEIN PACK, Model Number:DYNJ84239; 25) NEXTON PACK, Model Number:DYNJ88152; 26) PICC INSERTION TRAY W/O CATH, Model Number:PICCNC0004

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Identifiers

code_info
1) 00-401993O, UDI-DI: 10195327517366(each), 40195327517367(case), Lot Number: 23KBH528; 2) ACC010268B, UDI-DI: 1088994…1) 00-401993O, UDI-DI: 10195327517366(each), 40195327517367(case), Lot Number: 23KBH528; 2) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 21BDB379; 3) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 21CDB078; 4) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 21IDC222; 5) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 21JDB458; 6) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 22ADC105; 7) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 22BDA862; 8) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 22CDC433; 9) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 22DDA763; 10) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 22FDB218; 11) ACC010268B, UDI-DI: 10889942197477(each), 40889942197478(case), Lot Number: 23BDB271; 12) ACC010268B, UDI

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2094-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.