RecallRadar

FDA Device recall Z-2094-2023

BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201

On 2023-07-12 the FDA classified a Class II recall of BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201 by Carefusion 303, citing automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downt….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2094-2023
ClassificationClass II
Statusongoing
Initiated2023-05-22
Published2023-07-12
Last updated2026-09-13 11:56 UTC
CompanyCarefusion 303
DistributionUS

Product

BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191, 111-192, 111-193, 111-194, 111-195, 111-196, 111-197, 111-198; BD Pyxis CII Safe, Software: V8, 111-199-01, 111-200-01, 111-201-01, 111-202-01, 111-203-01, 111-204-01, 111-205-01, 111-206-01, 107-245-01, 107-246-01; BD Pyxis CII Safe, Software: V9, REF: 107-252-01, 111-215, 111-216, 111-219, 111-221. BD Pyxis CII Safe, Software: ES, REF: 107-254-01, 107-255-01, 1116-00, 1117-00; BD Pyxis CII Safe, Software: N/A, REF: 111-113, 111-114, 111-115, 111-160-201, 111-217, 111-218, 111-220, 111-222; Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025. Used with BD Pyxis CII Safe System: v7.x for WES7 User Guide, DIR 10000145747/ v8.1 User Guide, DIR 10000277635/v9.0 User Guide, DIR 10000354270/ ES User Guide v1.8, DIR 10000418517

Reason

Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downtime and/or potential electric shock.

Identifiers

code_infoUDI-DI: 10885403512520; Devices Manufactured May 04, 2006 - September 26, 2023

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2094-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.