RecallRadar

FDA Device recall Z-2093-2023

NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs i…

On 2023-07-12 the FDA classified a Class III recall of NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs i… by Neumodx Molecular, citing potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SA….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2093-2023
ClassificationClass III
Statusongoing
Initiated2023-05-15
Published2023-07-12
Last updated2026-09-13 11:56 UTC
CompanyNeumodx Molecular
DistributionUS

Product

NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs in transport medium and bronchoalveolar lavage (BAL) specimens from individuals suspected of COVID-19 by their healthcare provider (HCP). REF Number: 300800

Reason

Potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The instrument registers the non-human readable text as lot number 123910 from the barcode label.

Identifiers

code_infoGTIN: 10814278021264 Lot Number: 123909 Exp. Date: 08-Dec-2023

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2093-2023%22