RecallRadar

FDA Device recall Z-2091-2023

Aptio Automation Storage and Retrieval Module (SRM)

On 2023-07-12 the FDA classified a Class II recall of Aptio Automation Storage and Retrieval Module (SRM) by Siemens Healthcare Diagnostics, citing A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a dela….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2091-2023
ClassificationClass II
Statusongoing
Initiated2023-05-16
Published2023-07-12
Last updated2026-09-13 11:56 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Aptio Automation Storage and Retrieval Module (SRM)

Reason

A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.

Identifiers

code_infoModel No. 10713762 and 10715227; UDI-DI: 00630414596754; All units with software versions prior to 38.