FDA Device recall Z-2090-2012
Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryke…
- FDA Device
- Class II
- ongoing
- Published 2012-08-01
On 2012-08-01 the FDA classified a Class II recall of Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryke… by Stryker Howmedica Osteonics, citing stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one p….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2090-2012 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2012-04-30 |
| Published | 2012-08-01 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Howmedica Osteonics |
| Distribution | US |
Product
Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430. Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy, developed specifically for orthopaedic applications. Rejuvenate Modular Necks The Rejuvenate Modular Necks are made of Cobalt Chrome. The necks in this system consist of 4 lengths in 4 mm increments. They are 30 mm, 34 mm, 38 mm and 42 mm. They come in multiple neck angles and versions in each length. They are 127, 130 and 132. The versions are 0, 8, 16 degrees. In total there are 16 unique neck implants in the Rejuvenate Modular System.
Reason
Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an
Identifiers
| code_info | 510 K092561 all lots Rejuvenate (NLS/NLV) Rejuvenate (SPT) (4845-4XXX and NLS/NLV) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2090-2012%22