FDA Device recall Z-2089-2014
Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the…
- FDA Device
- Class II
- terminated
- Published 2014-07-30
On 2014-07-30 the FDA classified a Class II recall of Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the… by Perouse Medical, citing perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2089-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-05-10 |
| Published | 2014-07-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Perouse Medical |
| Distribution | n/a |
Product
Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter.
Reason
Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.
Identifiers
| code_info | Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 1203124…Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 12031249, 12031527, 12031248, 12041183, 12041186, 12031244, 12031461, 12031460, 12021404, 12031462, 12031245, 12021405, 12031405, 12031246, 12021401, 12041344, 12031237, 12031241, 12021403, 12041345, 12041184, 12031497, 12031240, 12021194, 12031524, 12031480, 12041181, 12031275, 12031242, 12021450, 12021449, 12041185, 12031525, 12031247, 12021402, 12021448. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2089-2014%22