FDA Device recall Z-2088-2017
GP321R
- FDA Device
- Class II
- terminated
- Published 2017-05-24
On 2017-05-24 the FDA classified a Class II recall of GP321R by Aesculap Implant Systems, citing diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction m….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2088-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-03-28 |
| Published | 2017-05-24 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Aesculap Implant Systems |
| Distribution | US |
Product
GP321R; ELAN 4 2-RING DIAMOND BURR COARSE D2.3; GP322R; ELAN 4 2-RING DIAMOND BURR COARSE D3.0; GP323R; ELAN 4 2-RING DIAMOND BURR COARSE D4.0; GP324R; ELAN 4 2-RING DIAMOND BURR COARSE D5.0; GP325R; ELAN 4 2-RING DIAMOND BURR COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Reason
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
Identifiers
| code_info | Model numbers: GP321R ; GP322R ; GP323R ; GP324R ; GP325R ; |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2088-2017%22