RecallRadar

FDA Device recall Z-2084-2025

Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: T…

On 2025-07-16 the FDA classified a Class I recall of Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: T… by Edwards Lifesciences, citing due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2084-2025
ClassificationClass I
Statusongoing
Initiated2025-05-14
Published2025-07-16
Last updated2026-09-13 11:57 UTC
CompanyEdwards Lifesciences
DistributionUS

Product

Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Reason

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

Identifiers

code_info
Lot Code: Model No: FEMII018A; UDI: 0069013031256; All lots up to: BSLC9148 Serial Numbers: 130468 130599 130704 130848…Lot Code: Model No: FEMII018A; UDI: 0069013031256; All lots up to: BSLC9148 Serial Numbers: 130468 130599 130704 130848 130953 131048 131286 131430 131626 131774 131775 132004 132005 132510 BSLC0025 BSLC0121 BSLC0422 BSLC1115 BSLC1470 BSLC4161 BSLC4472 BSLC5041 BSLC5293 BSLC5842 BSLC6163 BSLC6164 BSLC7336 BSLC7794 BSLC8660 BSLC9148

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2084-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.