FDA Device recall Z-2083-2015
The FLOW-i Anesthesia System Axillary O2 and Suction Module, Device Part number: 66 79 847 FLOW-i Anesthesia System: Is indicated for admin…
- FDA Device
- Class II
- terminated
- Published 2015-07-22
On 2015-07-22 the FDA classified a Class II recall of The FLOW-i Anesthesia System Axillary O2 and Suction Module, Device Part number: 66 79 847 FLOW-i Anesthesia System: Is indicated for admin… by Maquet Cardiovascular Us Sales, citing maquet has identified that in some FLOW-i Anesthesia System Auxiliary O2 and Suction Modules, the suctioning function cannot be activated or controlled by the suction unit on/off….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2083-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-03-02 |
| Published | 2015-07-22 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Maquet Cardiovascular Us Sales |
| Distribution | US |
Product
The FLOW-i Anesthesia System Axillary O2 and Suction Module, Device Part number: 66 79 847 FLOW-i Anesthesia System: Is indicated for administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.The system is intended for use in a wide range of population: from neonatal to adult patients. The system is intended for use in hospital environments, except for the MRI environment, by healthcare professionals trained in inhalation anesthesia administration. Auxiliary O2 and Suction Module: Is an option available for the FLOW-i Anesthesia System. The module contains two main parts: Flowmeter unit: The intended use of the flowmeter is to provide oxygen for patient therapy. Suction unit: The intended use is to extract body fluids from the stomach and airways.
Reason
Maquet has identified that in some FLOW-i Anesthesia System Auxiliary O2 and Suction Modules, the suctioning function cannot be activated or controlled by the suction unit on/off switch leading to a decrease or loss of suctioning capability of the modules.
Identifiers
| code_info | Device Part number: 66 79 847 Serial Numbers: 4, 14-24, 26-29, 31, 37, 40, 56, 138, 199, 205, 367, 369-372, 391, 394-3…Device Part number: 66 79 847 Serial Numbers: 4, 14-24, 26-29, 31, 37, 40, 56, 138, 199, 205, 367, 369-372, 391, 394-395, 416, 466-491, 517, 519, 525, 567-568, 570, 582-583, 606, 613, 619, 627, 629-631, 633, 635-641, and 643-665 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2083-2015%22