FDA Device recall Z-2082-2015

Autopen Classic Automatic insulin delivery pen, 3.0ml x 2 unit 2-42 units, AN3800, Lot 7JH, NDC 08470-3800-01, The product is a non-sterile…

On 2015-07-22 the FDA classified a Class II recall of Autopen Classic Automatic insulin delivery pen, 3.0ml x 2 unit 2-42 units, AN3800, Lot 7JH, NDC 08470-3800-01, The product is a non-sterile… by Owen Mumford Usa, citing the dials in the Autopens are not working - the dial does not stay dialed, but springs right back to 0. As a result, the patient could not dial up a dose of medicine to deliver vi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2082-2015
ClassificationClass II
Statusterminated
Initiated2014-12-11
Published2015-07-22
Last updated2026-09-13 11:53 UTC
CompanyOwen Mumford Usa
DistributionUS

Product

Autopen Classic Automatic insulin delivery pen, 3.0ml x 2 unit 2-42 units, AN3800, Lot 7JH, NDC 08470-3800-01, The product is a non-sterile automatic subcutaneous injection device which delivers insulin.

Reason

The dials in the Autopens are not working - the dial does not stay dialed, but springs right back to 0. As a result, the patient could not dial up a dose of medicine to deliver via an injection.

Identifiers

upc384703800014
code_infoLot 7JH, UPC Code 384703800014

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2082-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.