FDA Device recall Z-2074-2019
ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 05414734507585
- FDA Device
- Class I
- terminated
- Published 2019-08-14
On 2019-08-14 the FDA classified a Class I recall of ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 05414734507585 by St Jude Medical, citing electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely i….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2074-2019 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2019-06-20 |
| Published | 2019-08-14 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | St Jude Medical |
| Distribution | US |
Product
ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 05414734507585
Reason
Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
Identifiers
| code_info | Serial Number: 9836058, 9836059, 9836061, 9836062, 9836063, 9836064, 9836065, 9836066, 9836067, 9836219, 983…Serial Number: 9836058, 9836059, 9836061, 9836062, 9836063, 9836064, 9836065, 9836066, 9836067, 9836219, 9836220, 9836221, 9838312, 9838544, 9838546 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Other recalls by St Jude Medical
- CardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital System Model CM3100 -CardioMEMS Backend Web Application Model CM…
FDA Device · Class II · 2025-04-02 · St Jude Medical - Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model…
FDA Device · Class II · 2025-01-29 · St Jude Medical - CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Networ…
FDA Device · Class II · 2025-01-29 · St Jude Medical - CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Hosptial Electronics System (PES) REF CM3000 Merlin.net Patient Care Netwo…
FDA Device · Class II · 2025-01-29 · St Jude Medical - St. Jude Medical Agilis NxT Steerable Introducer, REF 408309
FDA Device · Class II · 2024-07-24 · St Jude Medical
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2074-2019%22