RecallRadar

FDA Device recall Z-2073-2012

ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS…

On 2012-08-01 the FDA classified a Class II recall of ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS… by Siemens Medical Solutions Usa, citing potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments.….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2073-2012
ClassificationClass II
Statusterminated
Initiated2012-06-25
Published2012-08-01
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionFL

Product

ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG. The ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Reason

Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.

Identifiers

code_infoARTISTE, Part #: 08139789, ONCOR Avant Garde, Part #: 05863472, ONCOR Impression Plus, Part #: 05857912, ONCOR Expression, Part #: 07360717

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2073-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.