FDA Device recall Z-2070-2025
The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 (or DM) Analyzer with plasma conversion that multip…
- FDA Device
- Class II
- ongoing
- Published 2025-07-09
On 2025-07-09 the FDA classified a Class II recall of The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 (or DM) Analyzer with plasma conversion that multip… by Hemocue Ab, citing glucose microcuvettes experienced transit time outside of limits, so stability through the whole microcuvette lifetime can no longer be guaranteed. The affected products may not h….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2070-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-05-12 |
| Published | 2025-07-09 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Hemocue Ab |
| Distribution | US |
Product
The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 (or DM) Analyzer with plasma conversion that multiplies the measured whole blood glucose value by a factor of 1.112 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. The box contains 4 vials with 25 microcuvettes in each vial.
Reason
Glucose microcuvettes experienced transit time outside of limits, so stability through the whole microcuvette lifetime can no longer be guaranteed. The affected products may not have been stored at the recommended labeled storage conditions which may have impacted the safety, quality, identity, potency, and purity of the product. This failure could increase the probability of an incorrect test result or a delayed test result.
Identifiers
| code_info | UDI-DI: 07311091107060, Lot: 2501708 (Only 4 boxes within specific McKesson Shipment) |
|---|