FDA Device recall Z-2070-2014
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set
- FDA Device
- Class II
- terminated
- Published 2014-07-23
On 2014-07-23 the FDA classified a Class II recall of Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set by Excelsior Medical, citing there is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2070-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-06-05 |
| Published | 2014-07-23 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Excelsior Medical |
| Distribution | US |
Product
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
Reason
There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.
Identifiers
| code_info | Product Code PA03 - Lot Nos.: 03030813, 03050813 and 03091213 Product Code BX03, Lot Nos.: 64031513, 64073113, 64081313, 64090513 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2070-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.