Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack…

On 2026-05-20 the FDA classified a Class I recall of Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack… by Windstone Medical Packaging, citing affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2066-2026
ClassificationClass I
Statusongoing
Initiated2026-04-02
Published2026-05-20
Last updated2026-09-13 11:58 UTC
CompanyWindstone Medical Packaging
DistributionUS

Product

Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F

Reason

Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.

Identifiers

code_infoPack AMS6908E UDI-DI B098AMS6908E0 Lot 214875 Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2066-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.