RecallRadar

FDA Device recall Z-2062-2018

ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use on ADVIA 1800, 2400 and XPT Chemistry Systems

On 2018-06-13 the FDA classified a Class II recall of ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use on ADVIA 1800, 2400 and XPT Chemistry Systems by Siemens Healthcare Diagnostics, citing urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2062-2018
ClassificationClass II
Statusterminated
Initiated2018-03-27
Published2018-06-13
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use on ADVIA 1800, 2400 and XPT Chemistry Systems

Reason

Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.

Identifiers

code_infoLot # Expiry UDI# 408143 2018-06-28 00630414517513 418290 2018-06 - 28 00630414517513

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2062-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.