FDA Device recall Z-2058-2019

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product…

On 2019-07-31 the FDA classified a Class II recall of Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product… by Zyno Medical, citing administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2058-2019
ClassificationClass II
Statusterminated
Initiated2019-05-29
Published2019-07-31
Last updated2026-09-13 11:55 UTC
CompanyZyno Medical
DistributionMA

Product

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072

Reason

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Identifiers

upc00814371020198
code_infoLot Numbers:18037128,18065375, 19015969 UDl/UPC/GTIN: 00814371020198

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2058-2019%22