RecallRadar

FDA Device recall Z-2057-2014

Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel…

On 2014-07-23 the FDA classified a Class II recall of Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel… by Remel, citing bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of Streptococcus pneumoniae.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2057-2014
ClassificationClass II
Statusterminated
Initiated2014-06-11
Published2014-07-23
Last updated2026-09-13 11:53 UTC
CompanyRemel
DistributionCO, IL, IN, NE, NJ, PA, SD, VA

Product

Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS.

Reason

Bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of Streptococcus pneumoniae.

Identifiers

code_infoSegments 1, 5, and 6 only of Lot 492571, Exp. 2015-10

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2057-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.