RecallRadar

FDA Device recall Z-2055-2017

Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397

On 2017-05-17 the FDA classified a Class II recall of Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397 by Greatbatch Medical, citing greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received repor….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2055-2017
ClassificationClass II
Statusterminated
Initiated2017-04-23
Published2017-05-17
Last updated2026-09-13 11:54 UTC
CompanyGreatbatch Medical
DistributionTX

Product

Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).

Reason

Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.

Identifiers

code_info
Lot numbers : Linear Straight Broach Handle (Rasp Handle): 2988946, 3138186, 3185601, 3287315, 3300879, 3333172, 343…Lot numbers : Linear Straight Broach Handle (Rasp Handle): 2988946, 3138186, 3185601, 3287315, 3300879, 3333172, 3433392, 3332697. EZ Clean" Linear Broach Handle: 3508491.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2055-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.