FDA Device recall Z-2054-2015

Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1. In vit…

On 2015-07-15 the FDA classified a Class II recall of Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1. In vit… by Siemens Healthcare Diagnostics, citing two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification. Issue #2: Timing issue in Vista 1500 may cause wrong rea….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2054-2015
ClassificationClass II
Statusterminated
Initiated2015-05-21
Published2015-07-15
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1. In vitro diagnostic analyzer.

Reason

Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification. Issue #2: Timing issue in Vista 1500 may cause wrong reagent or no reagent delivery. This may lead to unflagged, unexpected low results and results flagged with assay errors.

Identifiers

code_infoSiemens Material Number (SMN): 10284473, 10488224, 10444801, 10444802